Mol Premier® 1350 Pre-Heating Furnace

HIGH-TEMPERATURE FURNACE FOR PRECISE MEASUREMENT RESULTS

Mol Premier® 1350 Ausglühofen Detail

Precise measurement results

Mol Premier® 1350 Ausglühofen

Robust construction

Mol Premier® 1350 Ausglühofen Display

Easy operation

Mol Premier® 1350 Pre-Heating Furnace

The Mol Premier® 1350 Pre-Heating Furnace is a powerful and compact benchtop unit for the successful determination of elemental contents in the sensitive ppm range. For the determination of elemental contents in the sensitive ppm range, it may be necessary to reduce the blank values of the sample carriers such as combustion boats and crucibles.

The Mol Premier®1350 Pre-Heating Furnace meets or exceeds common standards such as ASTM E1019 “Standard Test Methods for Determination of Carbon, Sulfur, Nitrogen, and Oxygen in Steel, Iron, Nickel, and Cobalt Alloys by Various Combustion and Inert Gas Fusion Techniques” and ISO 9556 “Steel and iron – Determination of total carbon content – Infrared absorption method after combustion in an induction furnace” and ISO 4935 “Steel and iron – Determination of sulfur content – Infrared absorption method after combustion in an induction furnace“.

 The Mol Premier® 1350 Pre-Heating Furnace’s robust construction, easy operation and high temperature range of up to 1350 °C allow for extreme versatility of use. The laboratory furnace’s pre-heating tube, which is open on both ends, has a diameter of 29 mm and reaches temperatures of up to 1350 °C.

For the complete oxidation of samples at temperatures of up to 1550 °C, we recommend the Mol Premier® 1350 Combustion Furnace with a tube diameter of 28 mm.

For particularly large sample quantities, we recommend the Mol Premier® 1000 Combustion Furnace with a quartz glass continuous-flow reactor and a tube diameter of 50 mm.

Grafik mit Rakete und Pfeil

Operating Mode 1
Fast Heating

Direct control of the high-temperature furnace with maximum power – whenever things need to go fast.

Grafik mit Blatt und Pfeil

Operating Mode 2
Multi-Stage Heating

Through fully automatic, multi-stage heating, the set target temperature is reached – conserving resources for a long service life of the heating elements.

Grafik mit Uhr

Operating Mode 3
Timed Heating

The high-temperature furnace autonomously regulates the temperature according to a pre-set time program – energy-saving for low operating costs.

Heating Curve Mol Premier® 1350

Operating Mode 1 – Fast Heating

Heizkurve Mol Premier® 1350 Hochtemperaturofen

Product Features

Mol Premier® 1350 Pre-Heating Furnace
For pre-heating ceramic combustion boats and crucibles

Maximum temperature
1350 °C

Technology
Heating elements

Thermocouple
Type S according to DIN EN 60584

Combustion tube
Ceramic

Tube diameter
29 mm

Input voltage
230 VAC ± 10 %; 50/60 Hz, 16 A max.

Dimensions
approx. 33 x 57 x 55

Heated zone
15 cm

Average heat-up time
approx. 100 °C / min.

CPU
Powerful multimedia processor with four ARM cores (Cortex-A53, ARMv8)

Frequency
up to 1.4 GHz

Outputs (optional)
1 x full size HDMI output

Outputs (optional)
4 pole stereo output

Ethernet (optional)
Gigabit Ethernet over USB 2.0

Wi-Fi (optional)
Wi-Fi for IEEE 802.11b, g, n and ac in the 2.4 and 5 GHz range

Bluetooth (optional)
Bluetooth Classic und Low Energy (BLE)

USB (optional)
up to 4 x USB 2.0 Ports

Function Description

The high-temperature furnaces from the Mol Premier® seriesare resistance-heated, high-temperature furnaces.
Compared to conventional resistance heating wire, the heating elements used can reach sustained high temperatures of up to 1550 °C quickly and easily.

The Mol Premier® 1350 Pre-Heating Furnace reaches temperatures of up to 1350 °C.

The temperature can be regulated via a software-based PID controller. The electrical system is directly controlled via the analysis software, which can be operated with a capacitive touch display built into the front panel. The software displays the current temperature, allows the the target temperature to be set, regulates the temperature depending on the heating program used, and also limits the maximum current during the warm-up phase. The control value is then fed to the electrical system.

The current temperature is measured with a high-temperature thermocouple according to DIN EN 60584, amplified, digitized and made available to the analysis software. An additional sensor in the electrical system measures the ambient temperature and corrects the reference point of the thermocouple so that changes in the ambient temperature do not affect the furnace temperature.

The automatic current limitation prevents the maximum permissible output of the heating elements from being exceeded, thus enabling maximum service life of the heating elements used. The automatic current limitation functions independently and does not require manual readjustment over the course of the heating elements’ aging process.

Contact us for a consultation, no obligations

Images similar. Content may be subject to modifications or corrections.
Mol Premier® is a registered trademark of Mol Analytik GmbH.